The MQ-1601 is designed for measurements of the figure of merit, Q, of inductors and capacitors. Effective values of inductance, capacitance, resistance, distributed capacitance, and high frequency cables can be determined.
The overall frequency range is from 15.5kHz to 50MHz settable with use of a 6-digit counter display. |
| ■ |
Residual inductance is less than 10nH, an advantage in high Q measurements and for higher accuracy. |
| ■ |
Highly stable Q voltmeter circuit used for accurate measurements of low Q values. |
| ■ |
Pushbuttons used for frequency and Q ranging. |
| ■ |
6-digit counter for accurate frequency setting. |
| ■ |
ΔQ (delta Q) function provided for measuring small changes in Q; the scale expansion is ×10. |
| ■ |
Convenient Q LIMIT setting with lamp indication for quick sampling, checking and inspection. |
| ■ |
Correlation adjuster enables correlating results with other Q meters. |
| ■ |
Meter range indication with LEDs prevents mis-operation. |
| ■ |
Q measurements possible at 15.75kHz. |
| ■ |
With use of a special test jig (option), characteristics of dielectrics can be determined. |
| |
| Frequency Range |
15.5kHz to 50MHz in 7 bands |
| 15.5 to 50kHz, 50 to 160kHz, |
| 160 to 500kHz, 500 to 1600kHz, |
| 1.6 to 5MHz, 5 to 16MHz, |
| 16 to 50MHz |
| Frequency Counter |
6-digit, or 5-digit with final figure blanked. |
| Frequency Accuracy |
±(0.01% +1 count) |
|
| |
| Q Measurement |
5 to 750 in 7 ranges |
| 5 to 25, 10 to 40, 15 to 75, 25 to 120, |
| 50 to 250, 100 to 400, 150 to 750. |
| Q Accuracy |
±10% f.s. below 20MHz. |
| ±20% f.s. above 20MHz. |
| ΔQ Range |
0 to 7.5 (in Q=15 to 75 range). |
| 0 to 12 (in Q=25 to 120 range). |
| 0 to 25 (in Q=50 to 250 range). |
| 0 to 40 (in Q=100 to 400 range). |
| 0 to 75 (in Q=100 to 750 range). |
|
| |
| Tuning Capacitance |
19 to 480pF, overall |
Main: 22 to 480pF
(1pF divisions to 100pF and 5pF divisions above 100pF). |
| Vernier:-3 to +3pF |
| (minimum division=0.1pF) |
| Accuracy: ±(1%+1pF) |
|
| |
| Inductance Measurement |
0.1µ
H to 1H in 7 ranges at specified frequencies. |
| Accuracy |
±5% above 0.5µH. |
| ±10% below 0.5µH. |
|
| |
| Measuring Voltage |
7.5Vrms at Q=750: (Q×0.01)Vrms for 25/75/250/750 ranges. |
| (Q×0.00625)Vrms for 40/120/400 ranges. |
|
| |
| Q Meter Range Voltage |
Over±10% for correlation. |
|
| |
| Power Requirements |
AC 100V, 115V, 215V or 230V±10%
(voltage selector provided, 50/60Hz; approx. 20VA) |
| Dimensions, Overall |
Approx. 430(W) × 220(H) × 300(D)mm |
| Weight |
Approx. 10kg |
|
| * |
Specifications are subject to change without notice. |
|
|  |