category_09

MIA-1251 Impedance Analyzer download

mia1251

Impedance Analyzer MIA-1251 equips an accurate signal source whose frequency accuracy is ±;0.005%. In combination with wide test frequency range and 5 digits resolution, it provides accurate characteristic evaluation of passive components under the same conditions as actual use conditions of the D.U.T.
The great advantage of this series is Graphic Sweep Function; it graphs a parameter against frequency or AC drive level. Characterizing a parameter graphically helps user to understand how the passive component performs under different conditions.
Multi-Step Mode enables user to define test conditions up to 30 steps, and save/recall Max. 64 programs on a built-in memory. This function allows user to perform a series of characteristic evaluation which contains various measurement parameters easily.
Special software enables full remote control and collecting measurement data is attached as standard equipment.
The MIA-1251 can be widely used in not only production lines but also R&D purpose because of the high accurate test signal source and abundant measurement/indication functions.


FEATURES

Comprehensive measurement functions. Providing all frequently used measurement parameters.
  lZl, R, X, lYl,G,B,Cs,C,L,D,Q,θ, ESR, Rdc,Δ%,
Test Frequency Accuracy : ±0.005%
  Test Frequency Resolution : 5 digits
  Measurement Basic Accuracy : ±0.1%
Comparator Function compares measurements of D.U.T. with specified upper limits and lower limits of primary parameter, and displays the HIGH/PASS/LOW result in Analog Scale Indicator.  Either absolute measurement units or percentage limits are selectable.
In addition, Analog Scale Indicator helps user to intuitively know adjustment status of variable components.
Wide LCD display and soft key realize straight forward intuitive operation.

SPECIFICATIONS

 Measurement Item
Measurement Parameters lZl, R, X, lYl,G,B,Cs,C,L,D,Q,θ, ESR, Rdc,Δ%,
Measurement Circuit: Series or Parallel Mathematical Function:
Absolute and Percentage Deviation

 Test Signal
Test Frequency Range MIM-1250 : 20Hz to 500kHz 5 Digits Resolution
Test Frequency Accuracy ±0.005%
Output Impedance 100Ω
AC Test Signal Level 10mV to 2Vrms, 1mVrms steps
Measurement Parameters Measurement Range
 lZl, R, X 0.1mΩ to 100MΩ
 lYl, G, B 10nS to 1.000S
 C 0.01pF to 1F
 L 0.01nH to 100kH
 D 0.00001 to 9.9999
 Q 0.1 to 9999.9
 θ -180º to +180º
 Rdc 0.1mΩ to 100MΩ
 Δ% -999.99% to 999.99%
Basic Accuracy ±0.1% lZl, R, X, lYl, G, B, C, L
Measurement Speed DC Max: 30ms, Fast: 60ms, Medium: 120ms, Slow: 500ms
AC Max: 75ms, Fast: 150ms, Medium: 450ms, Slow: 600ms
LCD Display 320×240 Graphic Display
Calibration Mode Open Circuit Trimming and Short Circuit Trimming
Comparator Mode Primary Parameters, Absolute Measurement Units or Percentage
Comparator Indication HIGH/PASS/LOW and Beep
Measurement Ranging Auto or Manual
Trigger Mode Internal, Manual, External
Connection 4 Terminals BNC Guarded Connections

 Other Functions
Save and Recall Up to 64 Multi-Step programs can be stored and recalled.
Memory Back-Up Set-up conditions are saved when the power switch was turned off.
Interface RS-232C, GP-IB (Option), Hander = EXT I/O (Option)

 General Data
Power Requirement AC90 to 132V AC198 to 264V
Line Frequency 50/60Hz
Operating Temp. Range 10ºC to 40ºC
Guarantee Temp. Range 18ºC to 28ºC
Dimensions Approx. 340 (W) ×150 (H) ×525 (D) mm
Weight Approx. 6.5kg
Software Measurement software for PC

Test leads, fixtures, accessories, etc. are available.  Please refer to the accessory catalog or contact us

* Specifications are subject to change without notice.