Adopting space saving design, the MIM-1201 has performance which measures the same measurement parameters as Meguro’s higher grade LCR meters, and provides ±0.1% basic accuracy measurements of passive components despite the most compact model in our LCR Meters. Test frequency range which covers from 100Hz to 100kHz (in 388 steps) allows measurement under almost the same conditions as actual use conditions of the D.U.T.
In addition, this series equips Comparator Function and Deviation Mode which are convenient for screening process. Comparator Function displays the PASS/FAIL results of both primary and secondary parameter simultaneously. In Deviation Mode, this unit displays deviations from nominal values of both parameters.
Special software enables full remote control and collecting measurement data is attached as standard equipment.
The MIM-1201 is suitable for manufacturing environment and provides accurate four-terminal measurements. |
| ■ |
Comprehensive measurement functions. Providing all frequently used measurement parameters. |
| |
lZl,R,X,lYl,G,B,Cs,Cp,Ls,Lp,D,Q,Rdc,Δ%,θ |
| ■ |
Comparator Function compares measurements of D.U.T. with specified upper limits and lower limits of both primary parameter and secondary parameter simultaneously, and reports the PASS/FAIL result of the D.U.T. by the front panel LED indicators and beep. |
| ■ |
In Deviation Mode, either absolute measurement units or percentage limits are selectable. |
| ■ |
A switchable DC 2V Bias Voltage Function is equipped as a standard equipment. |
| ■ |
Up to 127 instrument set-up conditions files can be saved/recalled in a built-in memory |
| Measurement Item |
| Measurement Parameters |
lZl,R,X,lYl,G,B,Cs,Cp,Ls,Lp,D,Q,Rdc,Δ%,θ |
|
| Test Signal |
| Test Frequency Range |
100Hz to 100kHz in 388 steps |
| AC Test Signal Level |
50mV 100mV 250mV 500mV 1V |
| DC Test Signal Level |
0.5V |
| DC Bias Voltage |
2V, Switchable |
| Measurement Parameters |
Measurement Range |
| lZl R X |
0.01mΩ to 100.00MΩ |
| lYl G |
0.0001nS to 1000S |
| Cs Cp |
0.01pF to 1F |
| Ls Lp |
0.01uH to 100KH |
| D |
0.0001 to 9.9999 |
| Q |
0.1 to 9999.9 |
| Rdc |
0.01mΩ to 100.00MΩ |
| Δ% |
-999.99% to 999.99% |
| θ |
-180º to +180º |
| Measurement Speed |
Fast: 200ms, Medium: 300ms, Slow: 500ms |
| LCD Display |
240×64 Dot-matrix Display |
| Calibration Mode |
Open Circuit Trimming and Short Circuit Trimming |
| Comparator Mode |
Primary and Secondary Parameters, Absolute Measurement Units |
| Comparator Indication |
PASS/FAIL LED Indicator and Beep |
| Deviation Mode |
Absolute Measurement Units or Percentage limits Measurement Ranging Auto or Hold |
| Trigger Mode |
Internal, Manual, External |
| Connection |
4 Terminals BNC Guarded Connections |
|
| Other Functions |
| Save and Recall |
Up to 127 instrument set-up conditions files can be saved and recalled. |
| Memory Back-Up |
Set-up conditions are saved when the power switch was turned off. |
| Interface |
RS-232C, GP-IB [Option], Hander = EXT I/O [Option] |
|
| General Data |
| Power Requirement |
AC90 to 132V AC198 to 264V |
| Line Frequency |
50/60Hz |
| Operating Temp. Range |
10ºC to 40ºC |
| Guarantee Temp. Range |
18ºC to 28ºC |
| Dimensions |
Approx. 340 (W) ×100 (H) ×330 (D) mm |
| Weight |
Approx. 5kg |
| Software |
Measurement software for PC |
|
| * |